NAME

       vmifucalib - Generate IFU calibrations from a set of flat and arc lamp exposures.

SYNOPSIS

       esorex vmifucalib [OPTIONS] FILE.sof

DESCRIPTION

       This  recipe  produces  the  extraction mask, the wavelength calibration, and the relative
       transmission correction, from a set of flat field and one arc lamp exposures (IFU mode).

       Input files:

         DO category:                 Type:       Explanation:       Required:
         IFU_SCREEN_FLAT              Raw         Flat field            Y
         IFU_ARC_SPECTRUM             Raw         Arc spectrum          .

         LINE_CATALOG                 Calib       Arc line catalog      .

         MASTER_BIAS                  Calib       Master bias           Y
         IFU_IDENT                    Calib       Fiber identification  .

         CCD_TABLE                    Calib       Bad pixel table       .

       Output files:

         DO category:                 Data type:  Explanation:
         IFU_MASTER_SCREEN_FLAT       FITS image  Stack of all raw flat fields
         IFU_ARC_SPECTRUM_EXTRACTED   FITS image  Extracted arc spectra
         IFU_FLAT_SPECTRUM_EXTRACTED  FITS image  Extracted flat field spectra
         IFU_TRACE                    FITS table  Extraction mask
         IFU_IDS                      FITS table  Wavelength calibration
         IFU_TRANSMISSION             FITS table  Transmission correction

       An arc spectrum exposure may not be specified in input, but in  that  case  no  wavelength
       calibration  nor  fiber  relative  transmission  correction  will be computed, and further
       reduction of associated scientific data will not be possible; the only  products  in  this
       case will be the stack of all the input raw flat field exposures, and the extraction mask.
       If the arc spectrum exposure is specified in input, then  also  a  line  catalog  must  be
       given.  The  fiber identification table, if given, ensures that all spectra will be safely
       identified and assigned to the correct spatial position, but in  most  cases  it  will  be
       possible  to  reduce  the  data even without specifying this table in the input SOF. A CCD
       table must be specified only if a bad pixel cleaning is requested.

       For more details, please refer to the VIMOS Pipeline User´s Guide.

OPTIONS

       --AllowSingleFrames <bool>
              Flat fields combination method is ignored. (bool; default: True). The full name  of
              this      option      for      the      EsoRex      configuration      file      is
              vimos.Parameters.stacking.singleframes [default = True].

       --StackMethod <str>
              Flat fields combination method. (str; default: ´MinMax´). The  full  name  of  this
              option  for  the  EsoRex  configuration  file  is  vimos.Parameters.stacking.method
              [default = MinMax].

       --KSigmaLow <float>
              Low threshold for K-sigma clipping method. (float; default: 5.0). The full name  of
              this      option      for      the      EsoRex      configuration      file      is
              vimos.Parameters.stacking.ksigma.low [default = 5.0].

       --KSigmaHigh <float>
              High threshold for K-sigma clipping method. (float; default: 5.0). The full name of
              this      option      for      the      EsoRex      configuration      file      is
              vimos.Parameters.stacking.ksigma.high [default = 5.0].

       --MinRejection <int>
              Number of lowest rejected values for rejection method. (int; default: 0). The  full
              name    of    this    option    for    the    EsoRex    configuration    file    is
              vimos.Parameters.stacking.minmax.minimum [default = 0].

       --MaxRejection <int>
              Number of highest rejected values for rejection method. (int; default: 1). The full
              name    of    this    option    for    the    EsoRex    configuration    file    is
              vimos.Parameters.stacking.minmax.maximum [default = 1].

       --BiasMethod <str>
              Bias removal method. (str; default: ´Zmaster´). The full name of  this  option  for
              the  EsoRex  configuration file is vimos.Parameters.bias.removing.method [default =
              Zmaster].

       --CleanBadPixel <bool>
              Bad pixel correction on master flat field. (bool; default: False). The full name of
              this  option  for  the EsoRex configuration file is vimos.Parameters.badpixel.clean
              [default = False].

       --ApplyTransmission <bool>
              Apply transmission correction to extracted arc lamp spectra (bool; default:  True).
              The   full   name   of   this   option   for   the  EsoRex  configuration  file  is
              vimos.Parameters.apply.transmission [default = True].

       --MaxIdsRms <float>
              Maximum tolerated RMS of residuals in IDS fit (float; default: 2.0). The full  name
              of  this  option  for  the EsoRex configuration file is vimos.Parameters.ids.maxrms
              [default = 2.0].

       --LineIdent <str>
              Arc line identification method. (str; default: ´Blind´).  The  full  name  of  this
              option  for  the  EsoRex  configuration  file is vimos.Parameters.line.ident.method
              [default = Blind].

       --MaxTraceRejection <int>
              Maximum percentage of rejected positions in fiber spectra  tracing  (int;  default:
              33).   The  full  name  of  this  option  for  the  EsoRex  configuration  file  is
              vimos.Parameters.trace.failed [default = 33].

       --ComputeQC <bool>
              Compute QC1 parameters (bool; default: True). The full name of this option for  the
              EsoRex configuration file is vimos.Parameters.quality.enable [default = True].

       Note  that  it  is possible to create a configuration file containing these options, along
       with suitable default values. Please refer to the details provided by the 'esorex  --help'
       command.

SEE ALSO

       The  full  documentation  for the vimos pipeline can be downloaded as a PDF file using the
       following URL:

              ftp://ftp.eso.org/pub/dfs/pipelines/vimos/vimos-pipeline-manual-6.9.pdf

       An  overview  over  the  existing  ESO  pipelines  can  be   found   on   the   web   page
       https://www.eso.org/sci/software/pipelines/.

       Basic documentation about the EsoRex program can be found at the esorex (1) man page.

       It  is  possible  to  call  the  pipelines  from python using the python-cpl package.  See
       https://packages.python.org/python-cpl/index.html for further information.

       The  other  recipes  of  the  vimos  pipeline  are  vimos_ima_bias(7),  vimos_ima_dark(7),
       vimos_ima_det_noise(7),  vimos_ima_fringe(7), vimos_ima_science(7), vimos_ima_standard(7),
       vimos_ima_twilight_flat(7),    vmbias(7),    vmdark(7),     vmdet(7),     vmifucombine(7),
       vmifucombinecube(7),  vmifuscience(7),  vmifustandard(7),  vmmoscalib(7), vmmosscience(7),
       vmspphot(7)

VERSION

       vmifucalib 4.1.7

AUTHOR

       ESO VIMOS Pipeline Team and VIMOS Consortium <https://support.eso.org>

BUG REPORTS

       Please report any problems to  https://support.eso.org.  Alternatively,  you  may  send  a
       report to the ESO User Support Department <usd-help@eso.org>.

LICENSE

       This  file  is  part of the VIMOS Instrument Pipeline Copyright (C) 2019 European Southern
       Observatory

       This program is free software; you can redistribute it and/or modify it under the terms of
       the  GNU  General  Public  License  as  published  by the Free Software Foundation; either
       version 2 of the License, or (at your option) any later version.

       This program is distributed in the hope that it will be useful, but WITHOUT ANY  WARRANTY;
       without  even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE.
       See the GNU General Public License for more details.

       You should have received a copy of the GNU General Public License along with this program;
       if  not,  write to the Free Software Foundation, Inc., 59 Temple Place, Suite 330, Boston,
       MA  02111-1307  USA